Testing Performance
To test the threshold accuracy (logic analyzer)
Using the Modify up arrow on the function generator, increase offset voltage in
4 4
1-mV increments until all activity indicators for pod 1 show the channels at a logic
high. Record the function generator voltage in the performance test record.