Dissolved oxygen (D. O.)
130
8.3.2

Drift control (AutoRead)

The AutoRead function checks the stability of the measurement signal.
The stability has a considerable impact on the reproducibility of the
measured values.
When using a probe with integrated stirrer (StirrOx G), the
measurement with AutoRead can be started using the key on the
probe. As soon as a stable measured value is reached, the stirrer
switches itself off.
m
1
Use
to call up the Display mode.
2
Call up the oxygen measured variable with
3
Immerse the D. O. probe in the test sample.
4
Activate the AutoRead function with
h
The
display indicator appears on the display.
The measured value is frozen (hold function).
5
Start the measurement with AutoRead by pressing
key on the StirrOx G probe. The
If the StirrOx G probe is used:
When a stable signal is reached, an acoustic signal is heard
h
and
stops flashing.
If the StirrOx G probe is used: The
disappears.
6
If necessary, start the next measurement with AutoRead by
g
pressing
7
To terminate the AutoRead function: Press the
Note
In order to accept the current measured value without waiting for the
end of the AutoRead function, simply press
measuring module or the key on the StirrOx G probe.
h
3
or the key on the StirrOx G probe.
inoLab 740
m
.
a
.
g
or the
display indicator flashes.
flashes.
3
display indicator
a
key.
g
on the terminal or
ba75446e05