Calibrating Via A Comparison Measurement - wtw inoLab 740 Operating Manual

Hide thumbs Also See for inoLab 740:
Table of Contents

Advertisement

inoLab 740
ba75446e05
07/2010
8.2.3

Calibrating via a comparison measurement *

1
Connect the D. O. probe to the measuring instrument.
c
2
Press the
3
Immerse the D. O. probe in the calibrating solution.
g
4
Press the
The AutoRead measurement begins. When the measured
value is stable, the input field for the Fac. correction factor
opens.
5
Perform a comparison measurement, e. g. Winkler titration,
with the calibrating solution (determine the nominal value).
Leave the measuring instrument switched on until the nominal
value is determined.
6
Increase/reduce the correction factor with
until the displayed measured value corresponds to the nominal
value.
m
7
Use
or [ESC] to leave the input field and finish the
calibration.
8
Remove the D. O. probe from the calibrating solution.
* for the inoLab Oxi 740 and BSB/BOD 740 measuring modules