Resetting To Default Settings - wtw inoLab 740 Operating Manual

Hide thumbs Also See for inoLab 740:
Table of Contents

Advertisement

pH
pH reset
70
6.1.5

Resetting to default settings

You can reset (initialize) the settings for the pH measurement to the
default settings.
In a reset, the following functions for the pH measurement are reset to
the default settings:
Measuring mode
Asymmetry U
Slope S (k) in mV
Calibration procedures
Temperature, manual
With the reset pH function, all the settings for pH measurements are
reset to the default values.
1
Switch to the Menu mode.
46
2
Use
to select the main Cal menu.
3
Select the Reset pH menu item with
4
Confirm the selected menu item with
The default settings for pH measurements have been restored.
The Menu mode is terminated.
pH
in mV
0 mV
Asy
-59.16 mV
AutoCal TEC
25 °C
82
g
inoLab 740
.
.