BOD
Start measurement with
routine
136
15
Check the data on the display.
If the data is correct: Continue with step 17.
To change the data: Continue with step 16.
16
Select Yes in the With routine field with
The New dilution water entry field disappears.
17
Start the measurement with [Start]. The Samples window
opens.
18
Select a routine name with
19
Swap between the entry fields with [Next].
20
Select Yes or No in the Dilution water field with
If New dilution water No was selected: The Number entry
field disappears. The measurement of the blank value
solutions is skipped. The blank value solutions last
measured are used for the BOD determination.
If New dilution water Yes was selected: The Number entry
field appears.
21
Confirm the settings with [Start].
If New dilution water No was selected: The Start
measurement Sample window opens. Continue with
step 28.
For New dilution water Yes: The Start measurement Dilution
water window opens. Continue with step 22.