Limits Of Use Of The Product; Proper Use; Foreseeable Improper Use; Tbl. 2: Permissible Ambient Conditions - Pfeiffer Vacuum HIPACE 300 Operating Instructions Manual

Turbopump
Table of Contents

Advertisement

Available languages

Available languages

► Never put the device into operation with the high vacuum connection open.
► Keep lines and cables away from hot surfaces (> 70°C).
► Never fill or operate the unit with cleaning agents or cleaning agent residues.
► Do not carry out your own conversions or modifications on the unit.
► Observe the unit protection class prior to installation or operation in other environments.
2.4 Limits of use of the product
Installation location
Air pressure
Installation altitude
Rel. air humidity
Protection class
Overvoltage category
Permissible protection degree
Degree of pollution
Ambient temperature
Permissible surrounding magnetic field for air cooling
Maximum irradiated thermal output for air cooling
Maximum irradiated thermal output for water cooling
Maximum permissible rotor temperature of the turbopump
Maximum permissible bakeout temperature at the high
vacuum flange
Tbl. 2:
2.5 Proper use
► Use the turbopump only for generating vacuum.
► Use the turbopump only in combination with a suitable backing pump that can deliver up to the
► Use the turbopump only in closed indoor areas.
► Use the turbopump only for the evacuation of dry and inert gases.
2.6 Foreseeable improper use
Improper use of the product invalidates all warranty and liability claims. Any use that is counter to the
purpose of the product, whether intentional or unintentional, is regarded as improper use; in particular:
● Establishing the voltage supply without correct installation
● Installing with non-specified fastening material
Permissible ambient conditions
Notes on ambient conditions
The specified permissible ambient temperatures apply to operation of the turbopump at
maximum permissible backing pressure or at maximum gas throughput, depending on the
cooling type. The turbopump is intrinsically safe thanks to redundant temperature monitor-
ing.
● The reduction in backing pressure or gas throughput permits operation of the turbo-
pump at higher ambient temperatures.
● If the maximum permissible operating temperature of the turbopump is exceeded, the
electronic drive unit first reduces the drive output and then switches it off where neces-
sary.
required maximum fore-vacuum pressure.
weatherproof (internal space)
750 hPa to 1060 hPa
max. 5000 m
max. 80%, at T <31°C,
up to max. 50% at T <40°C
III
II
IP44,
Type 12 according to UL 50E
2
5 °C to 30 °C with convection cooling
without gas throughput
5°C to 35°C with air cooling
5°C to 40°C with water cooling
5.5 mT
2.4 W
1.2 W
90 ℃
120 ℃
Safety
71/116

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Hipace 800Hipace 350Hipace 400Hipace 450Hipace 700