Operation options
6.4
Post-connection check
Is the device undamaged (visual inspection)?
Does the supply voltage in the system match the data on the nameplate of the device?
Do the cables used meet the necessary specifications?
Are the maximum values for voltage and current observed at the pulse and status output?
Do the cables have adequate strain relief?
Is the cable type route completely isolated?
Without loops and cross-overs?
Are the power supply and signal cables correctly connected?
7
Operation options
7.1
Overview of operation options
7.1.1
Configuration
Customer-specific configuration with DeviceCare and FieldCare
The device is operated via the DeviceCare and FieldCare operating programs. DeviceCare and
FieldCare are universal service and configuration programs from Endress+Hauser. Connection
is by means of the service adapter with an FXA291 service interface.
More information on DeviceCare and FieldCare and the operation of these programs is
available in the online help.
DeviceCare and FieldCare can be used for the following:
• Configuration of device functions
• Data backup of device parameters
• Measuring-point documentation
28