These Brief Operating Instructions are not a substitute for the Operating Instructions pertaining to the device. Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer • Smartphone/tablet: Endress+Hauser Operations app...
Page 2
Dosimass Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
About this document Dosimass About this document Symbols 1.1.1 Safety symbols DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING This symbol alerts you to a potentially dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
Dosimass Safety instructions Symbol Meaning Potential equalization connection (PE: protective earth) Ground terminals that must be connected to ground prior to establishing any other connections. The ground terminals are located on the interior and exterior of the device: • Interior ground terminal: potential equalization is connected to the supply network.
Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
Dosimass Incoming acceptance and product identification Workplace safety When working on and with the device: ‣ Wear the required personal protective equipment as per national regulations. Operational safety Damage to the device! ‣ Operate the device in proper technical condition and fail-safe condition only.
• Enter the serial numbers from the nameplates in the Device Viewer (www.endress.com/deviceviewer): all the information about the device is displayed. • Enter the serial numbers from the nameplates into the Endress+Hauser Operations app or scan the DataMatrix code on the nameplate with the Endress+Hauser Operations app: all the information about the device is displayed.
Dosimass Storage and transport Transporting the product Transport the measuring instrument to the measuring point in the original packaging. Do not remove protective covers or caps installed on process connections. They prevent mechanical damage to the sealing surfaces and contamination in the measuring tube.
Mounting Dosimass Mounting Mounting requirements 5.1.1 Mounting position Installation point A0028772 Installation in down pipes However, the following installation suggestion allows for installation in an open vertical pipeline. Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter prevent the sensor running empty while measurement is in progress.
Page 11
Dosimass Mounting A0028773 2 Installation in a down pipe (e.g. for batching applications) Supply tank Sensor Orifice plate, pipe restriction Valve Filling vessel Ø orifice plate, pipe restriction [mm] [in] [mm] [in] ¹⁄₂₄ 0.03 ¹⁄₁₂ 0.06 ¹⁄₈ 0.12 ³⁄₈...
Page 12
Mounting Dosimass Recommended orientation for DN 1 to 4 (¹⁄₂₄ to ¹⁄₈ ") Orientation Recommendation Vertical orientation A0015591 Horizontal orientation, transmitter at A0015589 Horizontal orientation, transmitter at bottom A0015590 Horizontal orientation, transmitter at side A0015592 This orientation is recommended to ensure self-draining.
Page 13
Dosimass Mounting Orientation Recommendation Horizontal orientation, transmitter at bottom A0015590 Horizontal orientation, transmitter at side A0015592 This orientation is recommended to ensure self-draining. Applications with low process temperatures may reduce the ambient temperature. To maintain the minimum ambient temperature for the transmitter, this orientation is recommended.
Page 14
Mounting Dosimass A0003768 Measuring device Filling valve Vessel Filling systems The pipe system must be completely full to ensure optimum measurement. A0003795 4 Filling system Measuring device Filling valve Vessel Endress+Hauser...
Dosimass Mounting Inlet and outlet runs No special precautions need to be taken for fittings that create turbulence, such as valves, elbows or T-pieces, as long as no cavitation occurs → 15. A0029322 A0029323 For the dimensions and installation lengths of the device, see the "Technical Information"...
Page 16
Mounting Dosimass A0034391 5 Thermal insulation with exposed extended neck Heating NOTICE Electronics can overheat due to elevated ambient temperature! ‣ Observe maximum permitted ambient temperature for the transmitter. ‣ Depending on the medium temperature, take the device orientation requirements into account.
Page 17
Dosimass Mounting 5.1.3 Special mounting instructions Information for filling systems Correct measurement is only possible if the pipe is completely full. We therefore recommend that some test batches be carried out prior to production batching. Circular filling system A0003761 Tank...
Page 18
Mounting Dosimass Linear filling system A0003762 Tank Measuring instrument Filling valve Vessel Endress+Hauser...
Page 19
Sensor holder DN 1 to 4 (¹⁄₂₄ to ¹⁄₈") • The appropriate sensor holder must be used for all applications with increased safety or load requirements and for sensors with clamp process connections. • The Endress+Hauser sensor holder is generally recommended for mounting for all applications . 20...70 ( 0.79...2.75)
Page 20
Mounting Dosimass WARNING Strain on pipes! Excessive strain on an unsupported pipe can cause the pipe to break. ‣ Install the sensor in a sufficiently supported pipe. In addition to the use of the sensor holder, for maximum mechanical stability the sensor can also be supported on the inlet and outlet sides onsite at the installation location with the use of pipe clamps, for example.
All Dosimass measuring instruments are calibrated in accordance with state-of-the-art technology. Calibration takes place under reference conditions . Zero adjustment is therefore not required for the Dosimass as a general rule. ‣ Experience shows that a zero adjustment is advisable only in special cases.
Page 22
Mounting Dosimass Has the correct orientation for the sensor been selected → 11? • According to sensor type • According to medium temperature • According to medium properties (outgassing, with entrained solids) Does the arrow on the sensor nameplate match the direction of medium flow through the piping ?? ...
Dosimass Electrical connection Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
Electrical connection Dosimass Example: Terminated device plug with cable: Lumberg RKWTH 8-299/10 Total length of cable in the Modbus network > 50 m Use shielded twisted pair cable for RS485 applications. Example: • Cable: Belden item no. 9842 (for 4-wire version, the same cable can be used for the power supply) •...
Page 25
Dosimass Electrical connection RSE 8 M12x1 Out PFS1 Out PFS2 A0054873 6 Connection to device Coupling: Supply voltage, pulse/freq./switch output Connector: Supply voltage, pulse/freq./switch output PELV or SELV power supply 1 to Pin assignment Pin assignment Connection: Coupling (A) – Connector (B)
Page 27
Dosimass Electrical connection Pin assignment Connection: Coupling (A) – Connector (B) Assignment Supply voltage Service interface RX Service interface TX Supply voltage Not used – Pulse/frequency/switch output DQ – IO-Link communication signal C/Q – Service interface GND The pin assignment deviates from the IO-Link standard to enable compatibility with previous device versions and installations.
Page 28
Electrical connection Dosimass Version 1: Status input via connection A/B RSE 8 RSE 5 M12x1 M12x1 A0053319 8 Connection to device Coupling: Supply voltage, Modbus RS485, status input Connector: Supply voltage, Modbus RS485, status input Coupling: Switch output (batch)
Page 29
Dosimass Electrical connection Version 2: Status output via connection A/B RSE 8 RSE 5 M12x1 M12x1 A0053323 9 Connection to device Coupling: Supply voltage, Modbus RS485, status output Connector: Supply voltage, Modbus RS485, status output Coupling: Switch output (batch), status input...
Electrical connection Dosimass Pin assignment Connection: Coupling (A) – Connector (B) Connection: Coupling (C) – Connector (D) Assignment Assignment Supply voltage Status input Service interface RX Switch output (batch) 2 Service interface TX – Switch output (batch) 1 and 2,...
Page 31
Dosimass Electrical connection A0032652 Coupling Plug Device version: Modbus RS485, 2 batching outputs, 1 status output, 1 status input A0032534 A, C Coupling B, D Plug 6.3.2 Grounding Grounding is by means of a cable socket. Endress+Hauser...
Electrical connection Dosimass A0053306 Ensuring potential equalization No special measures for potential equalization are required. Ensuring the degree of protection The measuring device fulfills all the requirements for IP67 degree of protection, Type 4X enclosure. To guarantee IP67 degree of protection, Type 4X enclosure, carry out the following steps after the electrical connection: ‣...
Connecting the operating tool Using service adapter and Commubox FXA291 Operation and configuration can be performed using the Endress+Hauser FieldCare or DeviceCare service and configuration software. The device is connected to the USB port of the computer via the service adapter and Commubox FXA291.
Page 34
FieldCare Function range FDT-based (Field Device Technology) plant asset management tool from Endress+Hauser. It can configure all smart field units in a system and helps you manage them. By using the status information, it is also a simple but effective way of checking their status and condition.
DeviceCare Function range Tool for connecting and configuring Endress+Hauser field devices. The fastest way to configure Endress+Hauser field devices is with the dedicated "DeviceCare" tool. Together with the device type managers (DTMs) it presents a convenient, comprehensive solution. Innovation brochure IN01047S •...
System integration Dosimass System integration For detailed information on system integration, see the Operating Instructions for the device • Overview of device description files: • Current version data for the device • Operating tools • Compatibility with earlier model • Modbus RS485 information •...
Dosimass Diagnostic information Diagnostic information Faults are displayed on the home page of the DeviceCare and FieldCare operating tools once the connection to the measuring instrument has been established. Remedial measures are provided for each diagnostic event to ensure that problems can be rectified quickly.