Range Calibration (P650 To P654) - Siemens airanger DPL PLUS Instruction Manual

Table of Contents

Advertisement

P342 Start Ups
The number of times power has been applied since the Date Of Manufacture.
Primary Index
Values
Related

Range Calibration (P650 to P654)

There are two types of calibration possible:
Offset:
Sound Velocity: Adjusts speed of sound and changes the measurement calculations.
Do Offset calibration at any steady level unless a Sound Velocity calibration is also done.
If both calibrations are done then do Offset at a known high level and Sound Velocity at a
known low level.
P650 Offset Calibration
Calibrates Empty (P006) if the reported level is consistently high or low by a fixed amount.
Primary Index
Values
Related
Before using this feature, verify the following parameters are correct:
Empty (P006)
Temperature (P664)
Offset Reading (P062)
Offset Calibration
Begin with a steady level.
1.
Press TRANSDUCER
2.
Repeat Step One at least five times to verify repeatability.
3.
Measure the actual reading (use tape measure).
4.
Enter the actual value.
The deviation between the entered Empty (P006) value and the calibrated Empty value is
stored in Offset Correction (P652).
Page 58
Global
Range: 1 to 9999 (view only)
P340 Date of Manufacture
P342 Run Time
Adjusts the measurements by a fixed amount.
Transducer
Range: -999 to 9999
P006 Empty
P062 Offset Reading
to display the calculated reading.
AiRanger DPL Plus – INSTRUCTION MANUAL
P652 Offset Correction
P664 Temperature
7ML19981BD02