Installation Records (P340 to P342)

P340 (V) Date of Manufacture
View the date of manufacture of this DPS.
Values:
P341 (V) Run Time
View the accumulated number of days this DPS has been operating, since the
Date of Manufacture.
Values:
P342 (V) Start Ups
View the accumulated number of times power has been applied to the DPS
(following a power interruption), since the Date of Manufacture (P340).
Values:

Range Calibration Parameters (P650 to P654)

P650 Offset Calibration
Use this feature if the Empty (P006) value was estimated or if the reported sludge
level is consistently high or low by a fixed amount (e.g. 2 cm).
Before using this feature, verify the:
Empty (P006) value (as measured or estimated) was entered correctly.
Temperature (P664) is correct.
Offset Reading (P062) value (if used) was entered correctly.
To perform an Offset Calibration:
With the material at a high steady level...
1. Press
2. Repeat step 1 at least 5 times to overcome Echo Lock (P721) and verify
repeatability.
3. Measure the actual distance* (e.g. with a tape measure).
4. Enter the actual value. (The Offset Correction amount is stored in P652).
Values:
InterRanger DPS300
YY:MM:DD
0.000 to 9999
1 to 9999
to display the calculated distance*.
-999 to 9999
72
7ML19981BC01