Power
Note:
Before making the power connection, ensure proper voltage selection.
AC Supply Wiring
voltage switch
shown in the 'OFF'
position. Select
appropriate voltage.
Notes:
•
The equipment must be protected by a 15 A fuse or circuit breaker in the
building installation.
•
A circuit breaker or switch in the building installation, marked as the
disconnect switch, shall be in close proximity to the equipment and
within easy reach of the operator.
•
Never operate the AiRanger DPL Plus with the enclosure lid open, or
with the ground (earth) wire disconnected.
•
Ensure that any associated alarm or control equipment is disconnected
until satisfactory operation is verified.
DC Supply Wiring
Notes:
DC terminals shall be supplied from a SELV source in accordance with
IEC-1010-1 Annex H.
7ML19981BD02
33
34
35
36
37
24 V
!
BOARD B
AiRanger DPL Plus – INSTRUCTION MANUAL
34
100/115/200/230 V
50 / 60 Hz
Select voltage via switch.
33
34
35
36
37
24 V
L1
L2/N
35
36
37
Page 17