Universal, Linear (P050 = 9)

volume
level
This volume calculation creates a piece-wise linear approximation of the level/volume
curve. This option provides best results if the curve has sharp angles joining relatively
linear sections.
Enter a Level Breakpoint at each point where the level/volume curve bends sharply (2
minimum).
For combination curves (mostly linear but include 1 or more arcs), ensure numerous
breakpoints are entered along the arc, for best volume calculation accuracy.

Universal, Curved (P050 = 10)

This volume calculation creates a cubic spline approximation of the level/volume curve.
This option provides best results if the curve is non-linear, and there are no sharp angles.
volume
level
Select at least enough breakpoints from the curve to satisfy the following:
2 breakpoints very near the minimum level
1 breakpoint at the tangent points of each arc
1 breakpoint at each arc apex
2 breakpoints very near the maximum level
For combination curves, ensure at least 2 breakpoints are entered immediately before
and after any sharp angle (as well as 1 breakpoint exactly at the angle) on the curve.

Maximum Process Speed

The Maximum Process Speed (P003) to material level changes is designed to exceed the
most demanding installation requirements.