P810 Scope Displays
Captures echo profiles for display on an oscilloscope.
Primary Index
Values
Related
Use this feature to monitor the effects of Echo Processing changes.
Connect an oscilloscope to Display Board TP4 , TP5, and TP6.
Sweep = 10 us / div. to 1 ms / div. (x 100 for real time)
Amplitude = 1 V / div.
Trigger = external
Note:
Use of an analog oscilloscope is preferred over a digital oscilloscope.
Any combination of the following Scope Displays is available.
Display
Echo Profile
TVT Curve
Echo Marker
Echo Lock Window
Two methods to select Scope Displays:
Scrolling
1.
Press READING
2.
Press ARROWS
3.
Press ENTER
1/0 Values
Alternatively, a four-digit binary value may be entered, where a 0 turns the associated
signal display OFF, and a 1 turns the display ON.
1110 = PCn_ :
Echo Profile, TVT Curve, and Echo Marker displays ON
Echo Lock Window display OFF
7ML19981BD02
Transducer
Display: P , C, n, u, _
Preset: _ _ _ _ (display is off)
P832 TVT Shaper Adjust
P
to display the Auxiliary Function symbol.
to access the desired Reading display symbols.
with the desired display symbols displayed.
AiRanger DPL Plus – INSTRUCTION MANUAL
Symbol
C
n
u
Page 77