Profile Pointer (P817 To P825) - Siemens airanger DPL PLUS Instruction Manual

Table of Contents

Advertisement

Use the Scope Displays after pressing TRANSDUCER
parameter alterations. Take several measurements to verify repeatability and overcome
Echo Lock (P711) restrictions.
P816 Echo Time Raw
The time (in ms) from the transmit pulse to the processed echo.
Primary Index
Values

Profile Pointer (P817 to P825)

When one of these parameters is accessed, the scope display changes: The Echo Lock
Window becomes the Profile Pointer. The Profile Pointer may be moved to any point on
the Echo Profile to gain specific information about the profile according to the selected
parameter.
To move the Profile Pointer to a specific point, enter the desired value and it will move to
the nearest acceptable Echo Profile point.
Alternatively, to scroll the Profile Pointer along the Echo Profile:
1.
Press MODE
2.
Press
or
P
3.
When the Profile Pointer Parameters are exited and RUN mode is entered, the
scope display automatically changes back to include the Echo Lock Window.
P817 Profile Pointer Time
The time (in ms) from the transmit pulse to the Profile Pointer.
Primary Index
Values
Related
Page 78
Transducer
Range: 0.0 to 9999 (view only)
to display the Auxiliary Function symbol.
to move the Profile Pointer to the left or right respectively.
P
Transducer
Range: 0.000 to 9999 (view only)
P818 Profile Pointer Distance
P819 Profile Pointer Amplitude
P820 Algorithm
P821 Spike Filter
P822 Narrow Echo Filter
P823 Reform Echo
P825 Echo Marker Trigger