Connecting The Power Supply - Siemens SIPROCESS GA700 Compact Operating Instructions

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Available languages

Available languages

5.2.5.3

Connecting the power supply

Note
Securing the IEC connector
An IEC connector and a guard bracket are enclosed with the device. The guard bracket that
is screwed over the IEC connector protects the IEC connector from being pulled out
unintentionally.
Overview
Figure 5-8
Procedure
1. Connect the wires of the connecting cable to the IEC connector
power supply and PE conductor connections are correct!
Tighten the screws in the IEC connector with a torque of:
– 0.5 Nm at the terminals
– 0.3 Nm at the cable terminals (strain relief) and the connector enclosure.
2. Connect the appliance plug
Devices in explosion-proof design
Compact Operating Instructions, 11/2019, A5E35134047-04
Power supply unit
Equipotential bonding
Ethernet strain relief
Rear of the rack-mounted device, connecting power supply
Appliance plug
Guard bracket for the IEC connector
to the power supply unit
Connecting
5.2 Electrical connections
. Ensure that the
.
53