Connecting
6.2 Connecting SITRANS TDL
6.2.2

Power supply

Connection
Connect the power supply to HOST board P3 terminal. Use the customer cable provided with
the device or use a standard cable (shielded cable is preferred) with a maximum wire
diameter of 2.5 mm
and negative pole to the pin 26.
Protection
The device is protected by the eFuse with integrated reverse-input polarity protection. The
device can withstand positive and negative supply voltages up to ±60 VDC. In case of
overvoltage, undervoltage, overcurrent or reverse-input polarity the device is switched off
and the red diode on the HOST board (at the 24VDC section) is illuminated.
Power supply
The device does not have a power switch!
• Provide a protected circuit breaker in the vicinity of the device at the place of installation.
• The circuit breaker must be a switch or power switch, and must be labelled as the circuit
• At the supply point, the power supply must be isolated from dangerous live cables by
32
(14 AWG). Connect the positive pole of the power supply to the pin 25
2
WARNING
breaker for the device.
double or reinforced insulation in the case of devices with a 24 V supply voltage.
Compact Operating Instructions, 05\2023, A5E51659183-AA – revised 03/2024
SITRANS TDL