OPERATION
Close the lid of the centrifuge, now proceed as described above, to set all important run parameters.
If the lid isn´t closed, when storing the program in the display "rpm/rcf" (A-1), flashes alternately the
word "FirSt" and "CLOSE Lid" (see Figure 23). When starting the run without storing the program, in
the display "rpm/rcf" (A-1), flashes alternately the word ""First" and "PrESS StoreE", (see Figure 22).
Figure 21
Figure 22
For alteration of data, press the key "store" (12), for approx. 1 second. If the program is stored
correctly, the word "StorE" appears in the display "rpm/rcf" (A-1). As a result, the word "program"
(M10) disappears. As soon as the key "store" (12) is no longer displayed, the word "programm xx"
(M10) reappears. ( the xx stands for the chosen program place).
If all program numbers are occupied, take an old number that is not needed any longer and replace
it with the new parameters.

2.5.2 Recall of Stored Programs

To recall stored programs, press the key "prog" (11), with the lid already closed. Inside the display
"time" (A-3), appears "programm --"(M10). With the adjustable knob (1), pre-select the desired
program number.
In the respective displays, the stored values, for that program, will appear.
If there is not the correct rotor inside the centrifuge, for the pre-selected program, in the display
"rpm/rcf" (A-1) flashes the word "rotor" (M3). At the same time, the word "FALSE" and the stored
rotor number "22x. xx""(M4) flash in sequence of one another.
For all number marked text, please refer to Figure 23
Figure 23
20
Z216M_V2.21_eng
© Hermle Labortechnik GmbH