Pre-Selection Of Brake Intensity And Acceleration - Hermle Z 216 M User Manual

Table of Contents

Advertisement

OPERATION
- In the display, "time" (A-3) flashes the indication: "m : s" or "h : m", depending on the previous
setting.
To set the desired value, use the adjustable knob (1). After exceeding 59 min 50 sec, the indication
changes automatically to, "h : m". After exceeding 99 hours 59 min, the word "cont" appears in the
display, "time" (A-3). The continuous run can only be interrupted by pressing the key, "stop" (10).
The time counts down, as soon as the set speed is reached.
The display will always show the remaining running time, (see Figure 17)
For all number marked text, please refer to Figure 17
Figure 17

2.3.3 Pre-Selection of Brake Intensity and Acceleration

Selecting the key, "accel/decel" (5), this function is activated.
By pressing the key once, the word "accel" (M7) flashes, in the display "accel/decel" (A-2). The
desired acceleration can be pre-selected, with the adjustable knob (1). The value 0 is equivalent to
the lowest acceleration and the value 9 is equivalent to the highest acceleration.
By pressing the key "accel/decel" (5) twice, in the display "accel/decel" (A-2), indicates the word
"decel" (M8). Now the desired brake intensity can be pre-selected, with the adjustable knob (1). The
value 9 is equivalent to the shortest possible brake time and the value 0 to longest possible brake
time.
For all number marked text, please refer to Figure 18
See Table 4: "Acceleration and Deceleration Times", (APPENDIX P. IV). This table shows the
acceleration and deceleration times, for the acceleration and deceleration stages 0 to 9, for
permissible rotors.
Figure 18
18
Z216M_V2.21_eng
© Hermle Labortechnik GmbH

Advertisement

Table of Contents
loading

Table of Contents