Pre-Selection; Pre-Selection Of Speed / Rcf-Value; Pre-Selection Of Running Time - Hermle Z 216 M User Manual

Table of Contents

Advertisement

2.3 Pre-Selection

2.3.1 Pre-Selection of Speed / RCF-Value

Selecting the key, "rpm/rcf" (4), pre-selection is activated. By pressing the key once, the word "rpm"
(M5) flashes. By pressing the key again, the pre-selection of the centrifugal forces can be chosen.
The flashing word, "rcf" (M6), will appear. The desired values can be selected, with the adjustable
knob (1). In the display (A-1), the regulated value is shown permanently: before, during and after the
run.
For all number marked text, please refer to Figure 16.
Figure 16
As long as no rotor is inserted, the speed is adjustable between 200 rpm and the maximum
revolution of the centrifuge.
If there is a rotor in the centrifuge, the speed can only be pre-selected up to the maximum
permissible revolution of that rotor. It is the same with the pre-selection of the RCF-Value. The
setting range is between 20 xg and the maximum permissible centrifugal force of the rotor.
See Table 3: "Max. Speed and RCF-Values for Permissible Rotor", (see APPENDIX P. IV). All
important values are listed on this table.
ATTENTION:
Please check the maximum permissible revolutions of your test tubes! (Producer
Indication).

2.3.2 Pre-Selection of Running Time

The running time can be pre-selected in 3 different ranges: from 10 seconds up to 99 hours 59
minutes.
1. Range from: 10 seconds up to 59 minutes 50 seconds, in steps of 10 seconds
2. Range from: 1 hour up to 99 hours 59 minutes, in steps of 1 minutes
3. Range: Continuous Run "cont", can be interrupted by the key, "stop" (10).
-The running time can be pre-selected, with the lid opened or closed
-To activate the setting of the running time, press the key "time" (6).
© Hermle Labortechnik GmbH
Z216M_V2.21_eng
OPERATION
17

Advertisement

Table of Contents
loading

Table of Contents