OPERATING
If a program number is already occupied, in the display "rpm/rcf" (A-1), the words "rotor" (M3) and
e.g. "221.57" (M4) will appear, (see Figure 18). Free program numbers will appear as 0.
Figure 18
Close the lid of the centrifuge, now proceed as described above, to set all important run parameters.
If the lid isn´t closed, when storing the program in the display "rpm/rcf" (A-1), flashes alternately the
word "FirSt" and "CLOSE Lid" (see Figure 19). When starting the run without storing the program, in
the display "rpm/rcf" (A-1), flashes alternately the word "First" and "PrESS StoreE", (see Figure 20).
Figure 19
Figure 20
For alteration of data, press the key "store" (12), for approx. 1 second. If the program is stored
correctly, the word "StorE" appears in the display "rpm/rcf" (A-1). As a result, the word "program"
(M10) disappears. As soon as the key "store" (12) is no longer displayed, the word "programm xx"
(M10) reappears, (the xx stands for the chosen program place).
If all program numbers are occupied, take an old number that is not needed any longer and replace
it with the new parameters.

2.6.2 Recall of Stored Programs

To recall stored programs, press the key "prog" (11), with the lid already closed. Inside the display
"time" (A-3), appears "programm --"(M10). With the control field (1), pre-select the desired program
number. In the respective displays, the stored values, for that program, will appear.