Input Adjustment; Output Adjustment - Presys PCA-570-RM Technical Manual

Advanced pressure calibrator
Table of Contents

Advertisement

4.1. Input Adjustment

81
1) Input V, mA
Select the corresponding mnemonic and press the START button. Apply the signals corresponding to the
table below.
Note that the values applied should only be close to the values in the table.
After applying the signal, record the values at the 1st and 2nd calibration points (PNT1 and PNT2).

4.2. Output Adjustment

The adjustment of the outputs (except for Probe) is done in steps (STEPS). For each STEP the calibrator
generates a signal to be measured and recorded.
1) Output V, mA
For these output ranges the screen will show three buttons: STEP1, STEP2, and STEP3. Select STEP1.
The generated signal should be measured by a standard and the value should be saved by pressing the SAVE
button. Repeat the procedure for STEP2 and STEP3.
Input V
Single Range
Input mA
Single Range
PNT1
0,0000 V
PNT1
0,0000 mA
PNT2
11,0000 V
PNT2
20,0000 mA