Presys PCA-570-RM Technical Manual page 27

Advanced pressure calibrator
Table of Contents

Advertisement

The SWITCH input (CONTACT) count for measure the continuity of an external contact connected at the
input SWITCH on PCA-570. When there is continuity, the input show CLOSE, the opposite show OPEN.
23
By selecting previously the indication of pressure in the channel-2, in SWITCH is enable the function of
automated PRESSURE SWITCH test. On this function is possible verify the values of opening and closing contact at
pressure switch. For this, is necessary a pressure source that varies in a ramp more slowly, for high precision. For
this tests, is recommended increase the read rate of the A/D converter to FAST. See more information's in the
section 2.4.7 - A/D RATIO.
Select SWITCH and then PRESSURE SWITCH. The next screen is shown. Fill the camps of pressure
switch identification and the values to TRIP (value for closing/opening of pressure switch relé), DEAD ZONE (value of
pressure switch hysteresis) and TRIP ERROR/DEAD ZONE ERROR (values of tolerance for error of trip and dead
zone for pressure switch approval/disapproval). For issue of test report, use the function TASKS.
Fig. 14 - Pressure Switch Test