Presys PCA-570-RM Technical Manual page 67

Advanced pressure calibrator
Table of Contents

Advertisement

63
Fig. 46 - Task Information
When finalizing the task information, select the Preliminary/Final bar. Add each calibration point (point,
channel 2 of the calibrator) and its corresponding expected value (expected, channel 1 - calibrator input), both in the
As found calibration (calibration done before adjustment) and As left (calibration done after adjustment). The points
can also be generated with the help of the AUTO button. By pressing this button, simply enter the maximum and
minimum values of the calibration range and the number of points and a list of points will be generated considering
equal steps and a linear scale. Also fill in the number of repetitions (rep.) of the readings, and the calibration strategy
(start to end point ↑, end to start point ↓, etc.). By filling the REP field of the As found calibration with the value 0, the
task will only have the As left calibration. The REP field of the As left calibration must be filled in with the minimum
value 1.