Presys PCA-570-RM Technical Manual page 70

Advanced pressure calibrator
Table of Contents

Advertisement

66
The PCA-570 starts to perform the calibration automatically, generating the setpoints registered in the task
and taking the reading from the instrument to be calibrated. If NONE is selected for the calibrator input or output, the
calibrator prompts you to enter the data manually. The result is displayed on the screen, and a progress bar indicates
the remaining calibration time. At the end of the calibration, a report is presented with the values generated, the
values obtained, how much was expected, and the errors. If the error is above the value registered for the task, the
line appears in red.
The first time a task is executed, it is saved as As found (before adjustment). If it is run again, it will be
saved as As left (after adjustment). Tasks created with only As left are run only once. The results are saved in the
calibrator and can be viewed at any time in the RUN option.