Presys PCA-570-RM Technical Manual page 55

Advanced pressure calibrator
Table of Contents

Advertisement

For example, for a pressure transmitter, channel-2 must be set as Pressure Input and the pressure must be manually
generated (with a pump, for example) and connected to the PCA-570. To enable channel 2 as input, press the icon
51
and enable the CA-2 option. To return to the Calibrator screen press the BACK button. The channel-2
measurement or generation signal will act as the default value for the instrument's range setting.
Generate at the transmitter input the signal corresponding to the lower range value and press the button
The transmitter will generate 4 mA for this value. Generate the signal corresponding to the upper range value and
press the button
Another way to adjust with reference is to enter the HART option. Go back to the main menu by pressing the
icon
and the START button. Select the HART
CONNECT.
Fig. 36 - HART
. The transmitter will generate 20 mA for this value.
®
Fast Adjustment with Reference
®
option, set the connection type, address and then press
.