Input Adjustment - Presys PC-507-IS Technical Manual

Pressure calibrator
Table of Contents

Advertisement

Options for OUT adjustment are:
There is no order or interdependence between the adjustments.

4.1. Input Adjustment

Select the corresponding mnemonic and apply the signals presented in the
tables below.
When calibrating inputs, the display shows on the 2
by the calibrator and on the 1
Note that the applied signals just need to be close to the values shown in the
table.
Once the signal has been applied, store the values of the 1
adjustment points by pressing keys 1 (1
 V
st
line the same value is expressed as a percentage.
V Input
Single range
mA Input
Single range
mA
st
point) and 2 (2
st
1
point
2
0.00000 V
11.00000 V
st
1
point
2
0.0000 mA
20.0000 mA
nd
line the value measured
nd
point).
nd
point
nd
point
Page 50