Report Generator Mode
In the option to issue calibration reports, a more simplified document will be issued, with the calibration data,
62
both AS LEFT (final) and AS FOUND (preliminary), if any. It states whether the calibrated instrument is PASS/ FAIL
according to the registered acceptance criteria.

2.7.1. Creating Tasks

To generate tasks, from the main menu, select CALIBRATOR. Select the desired signal type on channel 1
and channel 2. For example, to calibrate a pressure transmitter, on the output enable CA-2 and select Pressure, and
on the input select mA (which will be connected to the current output of the transmitter). For a voltage indicator, for
example, select output V (which will be connected to the indicator's input) and input none (the instrument's display
must be entered manually).
Press the icon
Fill in, at least, the serial number of the instrument to be calibrated, TAG of the instrument, time to stabilize
each point (waiting time, in seconds). In Acceptance you can enable if the certificate will show a Pass/Fail report. In
this case, you must also fill in the maximum error allowed for the instrument to be calibrated (in % of the span,
reading or full scale). Filling the stabilization time with the value 0 (zero), the calibrator asks the operator to confirm
manually the acquisition of the point.
, and select TASKS > CREATE NEW TASK.