Presys TA-1200P Technical Manual

Temperature advanced calibrators
Table of Contents

Advertisement

Quick Links

®
Temperature Advanced
Temperature Advanced
Temperature Advanced
Calibrators
Calibrators
Calibrators
Calibrators
Calibrators
TA-1200P
TA-1200P
TA-1200P
TA-1200P
TA-1200P
Technical Manual
EM0295-03

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TA-1200P and is the answer not in the manual?

Questions and answers

Summary of Contents for Presys TA-1200P

  • Page 1 ® Temperature Advanced Temperature Advanced Temperature Advanced Calibrators Calibrators Calibrators Calibrators Calibrators TA-1200P TA-1200P TA-1200P TA-1200P TA-1200P Technical Manual EM0295-03...
  • Page 2 : Series TA-1200P Description Dry-Block Temperature Calibrator Block Temperature Calibrator Block Temperature Calibrator 2014/35/EC...
  • Page 4 The warranty conditions are available in our sites: The warranty conditions are available in our sites: The warranty conditions are available in our sites: www.presys.com.br/warranty www.presys.com.br/warranty www.presys.com.br/warranty...
  • Page 5: Table Of Contents

    PRESYS  Instrumentos e Sistemas TA-1200P Table of Contents 1 - Introduction........................1 ........1.1 - Technical Specifications...................2 ............2 2 1.1.1 - Input Technical Specifications................3 ..................3 3 1.1.2 - Special Software Features ................4 ....................4 4 1.2 - Order Code .......................4...
  • Page 6: Introduction

    RTD, TC and mA. the functions of dry block, standard thermometer and calibrator for RTD, TC and mA. • TA-1200P calibrator model generates temperatures from 50 ºC (122 ºF) to 1200 ºC TA- - 1200P calibrator model generate 1200P calibrator model generate (2192 ºF).
  • Page 7: Technical Specifications

    It has HART® communication for reading and setting parameters of field devices that have this protocol. TA-1200P has also many other features, such as: • RTD input for 2, 3 and 4 wires. Table IEC 60751, JIS or Callendar-Van Dusen RTD input for 2, 3 and 4 wires.
  • Page 8: Input Technical Specifications

    PRESYS  Instruments TA-1200P 1.1.1 - Input Technical Specifications Input Ranges Resolution Accuracy Remarks  0.01 % FS* millivolt -150 mV to 150 mV 0.001 mV > 10 M input  0.02 % FS 150 mV to 2450 mV 0.01 mV...
  • Page 9: Special Software Features

    PRESYS  Instruments TA-1200P 1.1.2 - Special Software Features Special Function: SCALE: makes the scaling of mA input. Memory Manager: stores configuration types predefined by the user. tores configuration types predefined by the user. tores configuration types predefined by the user.
  • Page 10: Accessories

    PRESYS  Instruments TA-1200P 1.3 - Accessories • Dry Block Insert: Order Code for Order Code for Order Code for Inserts Holes TA-1200P 1200P IN01 1 x 3/4" 06.04.0031-00 06.04.0031 06.04.0031- - 00 IN02 1 x 1/2" 06.04.0032-00 06.04.0032 06.04.0032- - 00 IN03 1 x 6.0mm and 3 x 1/4"...
  • Page 11: Initial Usage

    1. 1.5 5 - - Mounting the insert inside the furnace Mounting the insert inside the furnace In the core of the dry block TA-1200P there is a ceramic tube. Therefore, for In the core of the In the core of the dry block TA dry block TA safety purposes, the insert and the insulators are separated.
  • Page 12: Mounting The Insert Inside The Furnace

    PRESYS  Instruments TA-1200P Fig 2 - Schematic View for Insert Mounting Schema Schematic View for tic View for Insert Mounting Insert Mounting 1.6 - Instruction for use of the optional Black Body insert Instruction for use of the optional Black Body insert...
  • Page 13 Connect the terminals of the thermocouple type N to the auxiliary input side of the furnace TA-1200P and set the reading of the input to N type thermocouple 1200P and set the reading of the input to N type thermocouple 1200P and set the reading of the input to N type thermocouple (CJC internal).
  • Page 14 PRESYS  Instruments TA-1200P Fig 4 - Schematic view for mounting of the black body cavity Fig 4 Fig 4 - - Schematic view for mounting of the black body cavity Schematic view for mounting of the black body cavity...
  • Page 15: Parts Identification

    PRESYS  Instruments TA-1200P 1.7 - Parts Identification Fig. 05 - Parts Identification Page 10...
  • Page 16: Calibrator Operation

    VIDEOS - features videos made by Presys to assist in the use of the calibrator, and can VIDEOS VIDEOS - - features videos made by Presys to assist in the use of the calibrator, and can features videos made by Presys to assist in the use of the calibrator, and can also store videos made by the user, see section 2.5.
  • Page 17: Calibrator Menu

    PRESYS  Instruments TA-1200P 2.1 - Calibrator Menu To select the probe set point or electrical input functions, from the main menu, press the CALIBRATOR button. The following screen is displayed. Fig. 07 - Calibrator Function Calibrator Function Calibrator Function At the top is shown the probe settings and values.
  • Page 18 PRESYS  Instruments TA-1200P In the STEP function, a step value can be configured, and the steps can be changed through the up and down arrows. In REFERENCE menu, you can configure the type of probe reference (see section 2.1.1 – Probe Reference). The chosen reference appears just below the...
  • Page 19: Probe Reference

    PRESYS  Instruments TA-1200P 2.1.1 – Probe Reference There are two different references to the thermal block: Internal Reference and External Reference. The Internal Reference is a sensor built into the block. The External Reference is an option for more accurate measurements. The is an option for more accurate measurements.
  • Page 20 PRESYS  Instruments TA-1200P Fig. 11 - Adding a new Reference Sensor Adding a new Reference Sensor Adding a new Reference Sensor After filling the blanks, click on SAVE button and confirm. The new sensor is now SAVE SAVE button and confirm. The new sensor is now button and confirm.
  • Page 21: Input Settings

    PRESYS  Instruments TA-1200P 2.1.2 - Input Settings The INPUT menu has the following options: Fig. 13 - Input Menu Options 3 3 - - Input Menu Options Input Menu Options For OHM measurement, you should also select between 2, 3 or 4 wires options.
  • Page 22 PRESYS  Instruments TA-1200P To create new CVD coefficients, press (edit), and ADD button. The curves appear in the created list as identified in ID. It must be filled the following blanks: ID: Set an identification for the sensor R0 (): The last resistance measurement in 0 °C of the sensor A, B, C: Callendar-Van Dusen coefficients Low (°C): Lower value of the reference sensor calibrated range...
  • Page 23 PRESYS  Instruments TA-1200P When the input sensor breaking occur (RTD, resistance or probe) the display will show the burn-out warning identified by question marks illustrated below: Fig. 16 - Burn-out Warning If an out of range signal is injected, a message of UNDER or OVER range...
  • Page 24 PRESYS  Instruments TA-1200P 2.1.2.1 - Input Connections Diagrams Fig. 17 - Input Connections Page 19...
  • Page 25: Special Function

    PRESYS  Instruments TA-1200P 2.1.3 - Special Function SCALE: For the current input, it is possible to use the scale function: Fig. 18 - Option for mA Input: SCALE Option for mA Input: SCALE Option for mA Input: SCALE It establishes a linear relationship between the mA input signal and what is shown...
  • Page 26: Saving Current Configuration (Memory Manager)

    PRESYS  Instruments TA-1200P Fig. 20 - Scale Function Configuration Scale Function Configuration Scale Function Configuration Note: To enable the Scale Function, turn the function ON before touching To enable the Scale Function, t To enable the Scale Function, turn the functio...
  • Page 27: Hart Configuration

    PRESYS  Instruments TA-1200P ® 2.2 - Hart Configuration The TA Series Calibrators can be used to read and set parameters in devices that ® ® have HART Communication Protocol. HART Protocol allows digital communication between master (in this case, the TA Calibrator) and the slave (field...
  • Page 28: Starting Communication

    PRESYS  Instruments TA-1200P Fig. 22 - Transmitter Powered by an External Power Supply Transmitter Transmitter P P owered by an owered by an E E xternal xternal P P ower ower S S upply upply ® ® ® mA INPUT + HART...
  • Page 29: Adjusting The Measurement Range Of A Hart Adjusting The Measurement Range Of A Hart

    PRESYS  Instruments TA-1200P On this screen you can also edit the unit of the primary variable and the input filter (damping). ® ® ® Fig. 23 - Adjusting the measuring range of the HART Adjusting the measuring range of the HART...
  • Page 30: Checking / Adjusting Hart Checking / Adjusting Hart

    PRESYS  Instruments TA-1200P Another way to do this adjust is entering in the HART option through the MAIN MENU, set the connection type, address and then press CONNECT. Select the MONITOR tab. In this screen are shown the value of the primary variable (PV) read by ®...
  • Page 31: Automatic Tasks

    PRESYS  Instruments TA-1200P ® Fig. 26 - Checking / Adjusting the mA HART Transmitter Output ransmitter ransmitter O O utput utput 2.3 - Automatic Tasks In TA Calibrators, can be created and performed automatic calibration tasks. This performed performed automatic automatic calibration tasks calibration tasks.
  • Page 32 PRESYS  Instruments TA-1200P Fig. 27 - Task Information T T ask Information ask Information Go to the As Found/ As Left tab. Add each point to be generated by the As Found/ As Left As Found/ As Left tab.
  • Page 33 PRESYS  Instruments TA-1200P tour the range and find the value of opening and closing the thermoswitch. The minimum value for this field is 300 s. Tip: If the Trip point and deadband are not known, try the THERMOSWITCH TEST to find an approximate value before creating the task.
  • Page 34: Performing Tasks

    PRESYS  Instruments TA-1200P 2.3.2 - Performing Tasks To perform a task created from the main menu select TASKS > EXPLORE TASKS. A list identifying the created work orders that have not been orders that have not been orders that have not been performed yet (●...
  • Page 35 PRESYS  Instruments TA-1200P Fig. 32 - Task Results ask R R esults esults The task data is saved in a PDF file in the internal memory card of the Calibrator The task data is saved in a PDF file in the internal memory card of the Calibrator The task data is saved in a PDF file in the internal memory card of the Calibrator and can be accessed connecting the calibrator to a computer.
  • Page 36: Data-Logger

    PRESYS  Instruments TA-1200P 2.4 - Data-Logger The TA Series Calibrators allow you to record series of measurements over time to display data in chart or table format. Select CALIBRATOR from the main menu and select the desired configuration for...
  • Page 37: Videos 2.5 - Videos

    PRESYS  Instruments TA-1200P Recording can also be programmed to start at a certain date and time in the LOGGER option. Just set the start time and end time of recording. During the defined range, the measured points are saved in an internal file in TA Calibrator.
  • Page 38 PRESYS  Instruments TA-1200P In this option is possible to identify your calibrator, choosing a TAG name, Owner name and Location. • Sound Options Press + or - to configure a value for the system audio volume. • Printer Config.
  • Page 39: Network 2.6.2 - Network

    PRESYS  Instruments TA-1200P 2.6.2 - Network In NETWORK tab you can configure the IP address for the Ethernet communication with the computer. The IP address can be dynamically configured communication with the computer. The IP address can be dynamically configured communication with the computer.
  • Page 40 PRESYS  Instruments TA-1200P Fig. 35 - TA Calibrator Web Server Fig. Fig. 35 35 - - TA Calibrator Web Server TA Calibrator Web Server In the Web Server, you can monitor the calibrator screen, change the setpoint and , you can monitor the calibrator screen, change the setpoint and , you can monitor the calibrator screen, change the setpoint and see the auxiliary input readings.
  • Page 41: 3 - - Safety Instructions 3 - Safety Instructions

    4 - Recommendations as regards Accuracy of Measurements Recommendations as regards Accuracy of Measurements Recommendations as regards Accuracy of Measurements PRESYS Temperature Advanced Calibrators are instruments of high accuracy ced C C alibrators are instruments of high accuracy alibrators are instruments of high accuracy...
  • Page 42: Calibration (Adjustment)

    PRESYS  Instruments TA-1200P 5 - Calibration (Adjustment) Warning: Enter the following options only after understanding them completely. Warning: Enter the following options only after understanding them completely. Warning: Enter the following options only after understanding them completely. Otherwise, it may be necessary to return the instrument to the factory...
  • Page 43: Probe Calibration

    PRESYS  Instruments TA-1200P resistors Point 1 Point 2 20.000  50.000  OHM3 100.000  500.000  OHM2 500.000  2200.000  OHM1 The cold junction calibration (Thermocouple) is performed measuring the mV(-) ) is performed measuring the mV( ) is performed measuring the mV(- - ) ) terminal temperature.
  • Page 44: Pid Control Parameters

    PRESYS  Instruments TA-1200P 5.3 - PID Control Parameters The TA-1200P temperature calibrator has a PID control algorithm to calculate the a PID control algorithm to calculate the a PID control algorithm to calculate the block control output. The dry block stability and response time features are related to the PID...

Table of Contents