identification describe in ID. Select SETTINGS for insert one identification or TAG for sensor in ID and inform the
range of using in MIN and MAX. Select PARAMETERS for insert the sensor curve, usually stated on the calibration
20
certificate of the same.
For Callendar-Van Dusen: Fill the value of R0 (Resistance at 0 °C) and the parameters A, B and C along
with its exponents.
For ITS-90: Fill the value of RPTW (Resistance at 0,01 °C) and select the range for negative and/or positive
curve before insert the parameters along with its exponents.
Fig. 12 - Configurable the Callendar-Van Dusen coefficients.