2.4.6. Save Actual Configuration (Memory Manager)

40
The PCA-570 calibrator supports several special functions that may become frequently used. In these
situations it is useful to store such settings in the instrument in order to save time.
After setting the calibrator the way you want (input type, output type, special functions), press the icon
MORE OPTIONS, and the button MEMÓRY MANAGER. At the bottom of the screen you can give this setting a
name and a description. Presses the button SAVE.
The operation that was being performed by the PCA-570 is now stored in the memory identified by its name.
To return to this configuration, even after turning the calibrator off and on, enter the MEMORY MANAGER option and
select the desired configuration name and press the LOAD button.
The SET-STANDARD button sets the actual calibrator configuration as the default setting. Thus, each time
the PCA-570 is turned on, this will be the initial calibrator setting.

2.4.7. AD/Ratio

The PCA-570 multi calibrator allows you to change the conversion rate of the Analog-to-Digital Converter. At
faster rates, the readings displayed in gauge mode are updated more frequently, and at slower rates, the calibrator
takes more readings and updates the values at a lower frequency.
For tests where reading speed is important, such as pressure switch and thermostat tests, it is
recommended to use a faster rate that can "capture" the value closest to relay open/close. For calibrations and other
tests, it is recommended to use a slower conversion rate, which gives more accurate readings.
To change the conversion rate, press the icon
between FAST, MEDIUM and SLOW
> MORE OPTIONS, and the button AD/RATIO. Select
.
>