For the CH option, you must enter the ADDRESS of the instrument you wish to communicate with and press the
CONNECT button. If the address of the instrument is not known, you can press the search button
49
search for instruments in the address range 0 to 15. For the FH option the device is found automatically.
Up to 15 instruments are allowed in a HART
instrument with address 0, in the mA INPUT + HART
to 20 mA) or digitally (HART
®
(HART
ONLY).
2.5.3. HART® Transmitter Measurement Range Adjustment (CH Option)
While the instrument is being connected, the DESCRIPTION tab shows information about the identification
of the instrument, such as TAG, manufacturer, description, message, date, measurement range and input filter
(damping). Some of these parameters can be changed in the STANDARD SETUP.
On the DESCRIPTION tab, the MIN and MAX fields indicate the measurement range of the HART
transmitter. For PV (primary variable) equal to the MIN value, the transmitter should generate 4 mA. For PV (primary
variable) equal to the MAX value, the transmitter should generate 20 mA. The maximum allowable range of the
transmitter is shown above (RANGE...). To edit the working range of the transmitter, simply change the MAX and
MIN values and press the SAVE RANGE button
In this screen you can also edit the unit of the primary variable and the input filter (damping).
®
®
®
). In the network connection, the only way to read the primary variable is digitally
network (addresses 1 to 15). In a connection with a single field
connection, the primary variable can be read either analog (4
, which will
®