Operating Range
Power Supply
Resolution
Display Accuracy
Stability
Power Consumption
Stabilization Time
Heating Rate
Cooling Rate
Calibration Volume
Radial Uniformity
Axial Uniformity (40 mm):
Axial Uniformity (100 mm):
Dimensions (H,W,D)
Weight
Warranty

1.1.1 - Input Technical Specifications

Ranges
millivolt
0 to 70 mV
TC-J
-210 to 1200 ºC / -346 to 2192 ºF
TC-B
50 to 250 ºC / 122 to 482 ºF
250 to 500 ºC / 482 to 932 ºF
500 to 1200 ºC / 932 to 2192 ºF
1200 to 1820 ºC / 2192 to 3308 ºF
TC-S
-50 to 300 ºC / -58 to 572 ºF
300 to 1760 ºC / 572 to 3200 ºF
TC-L
-200 to 900 ºC / -328 to 1652 ºF
TC-C
0 to 1500 ºC / 32 to 2732 ºF
1500 to 2320 ºC / 2732 to 4208 ºF
TC-Au-Pt
0 to 500 °C/ 32 to 932 °F
500 to 1000 °C/ 932 to 1832 °F
FS = Full Scale
TA-1200PLAB
50 to 1200 °C (122 to 2192 °F)
230 Vac 50/60Hz
0.01 ºC or 0.01 ºF
± 2.5 °C
± 0.1 ºC
2500 W
1 h
± 6 ºC / min
10 h (1100 ºC to 200 ºC)
∅ 34 mm / 300 mm depth (immersion)
± 0.1 ºC (for metallic insert)
± 0.20 full range
± 0.35 full range
Horizontal Model:
600 mm x 470 mm x 450 mm
43.0 kg
1 year
Resolution
Accuracy
0.0001 mV
± 0.005 % FS
0.01 ºC / 0.01 ºF
± 0.10 ºC / ± 0.20 ºF
0.01 ºC / 0.01 ºF
± 1.25 ºC / ± 2.50 ºF
0.01 ºC / 0.01 ºF
± 0.75 ºC / ± 1.50 ºF
0.01 ºC / 0.01 ºF
± 0.50 ºC / ± 1.00 ºF
0.01 ºC / 0.01 ºF
± 0.35 ºC / ± 0.70 ºF
0.01 ºC / 0.01 ºF
± 0.50 ºC / ± 1.00 ºF
0.01 ºC / 0.01 ºF
± 0.35 ºC / ± 0.70 ºF
0.01 ºC / 0.01 ºF
± 0.10 ºC / ± 0.20 ºF
0.01 ºC / 0.01 ºF
± 0.25 ºC / ± 0.50 ºF
0.01 ºC / 0.01 ºF
± 0.35 ºC / ± 0.70 ºF
ºC / 0.01 ºF
± 0.09 ºC / ± 0.12 ºF
0.01 ºC / 0.01 ºF
± 0.06 ºC / ± 0.12 ºF
TA-1200PLAB
Remarks
R
> 10 MΩ
input
IEC-60584
IEC-60584
IEC-60584
DIN-43710
W5Re / W26Re
W5Re / W26Re
ASTM E1751