System - Presys PCA-570-RM Technical Manual

Advanced pressure calibrator
Table of Contents

Advertisement

3.4. System

78
In the SYSTEM tab you can configure the calibrator volume, touch screen setting, calibrator identification,
language, printer, and security options.
To adjust the screen, press TOUCH SCREEN OPTIONS. Press the center of the + signs on the screen (use
of the touch screen stylus is recommended). After calibration, press the screen again at any point. Confirm the
adjustment and return to the SYSTEM screen.
Select the desired language and confirm with OK. The system must be restarted to save the new
configuration.
In this option you can identify the calibrator by choosing a TAG, company name (OWNER), and location.
Press + or - to set a value for the audio volume.
Initially, the instrument does not have a password. This setting can be changed in SECURITY OPTIONS.
To create a new user, press the key icon
press CREATE. You can add a signature to be used when issuing the TASKS function reports and certificates. To do
so, select the user and press SIGN.
Touch Screen Options
Language Setting
Calibrator Identification
Sound Options
Security Options
and then the user icon
. Select NEW, fill in the blanks and