information about Input and Output settings and the IP address. Press Back to return to calibrator mode or HOME for
the Main Screen.
18
Fig. 10 - Fast Browser and Secondary Menu
The channel 1 can only be configurable as input. The channel 2 can be configurable as output (default
setting) or input. The channel 2 allows PRESSURE and PROBE measurement. An example of using channel 2 as
input is a calibration of the pressure transmiter, when you want measure the transmiter current (input mA in CA-1)
and pressure also (input PRESSURE in CA-2). For use the channel 2 as input, press the icon
and enable the
option CA-2. For use the channel 2 as output, desable the option CA-2. For back Main Screen of the Calibrator press
the button BACK. For go to Main Screen, press the icon
, and the button HOME.