Table of Contents

Advertisement

3.1. Input adjustment (IN)

50
Select the corresponding mnemonic and apply the signals presented in the tables below.
When adjusting inputs, the display shows on the 2
same value is expressed as a percentage.
Note that the applied signals just need to be close to values shown in the table.
Once the signal has been applied, store the values of the 1
(1
point) and 2 (2
st
Input adjustment for  is performed in two steps:
point).
nd
mV Input
G_4
G_3
G_2
G_1
mV Input
Faixa única
mA Input
Faixa única
nd
line the value measured by and on the first line the
st
and 2
1
st
point
0.000 mV
0.000 mV
0.000 mV
600.00 mV
1
st
point
0.0000 V
1
point
st
0.0000 mA
nd
calibration points by pressing keys 1
2
nd
point
70.000 mV
120.000 mV
600.00 mV
2400.00 mV
2
nd
point
11.0000 V
2
point
nd
20.0000 mA