How To Measure Current Dc With The Module Zed7 01-Abx - Ahlborn ZED7 00-ABx Instruction Manual

Measuring module for voltage, current and power
Table of Contents

Advertisement

5 How to run a Measurement with the Measuring Module
4. Connect the voltage source to the sockets of the measuring module by
using the connecting cables whereby the red connecting cable must be
connected to the positive pole and the red socket.
5. Switch on the voltage at the voltage source.
➢ The measured values will be displayed on the ALMEMO
5.2 How to measure Current DC with the
Module ZED7 01-ABx
The device internal current measurement path is not intended for
continuous operation for currents greater than 10A. In that case, the
measurement can only be run for a maximum of 10 minutes. After this
period of time, the device must be cooled down to ambient
temperature.
Cable the system (especially for measuring voltages greater than 50V)
in power-off state. Only switch on the voltage after the cabling has
been set up completely.
To avoid electric shocks, do not touch any exposed parts or
connections during operation.
Use the supplied touch proof connector cables.
With no consumer present do not connect the current measurement
module directly to the voltage source. Possible resulting risks are
overheating or destruction of the module.
1. Connect a consumer to the current source.
- 16 –
®
measuring device.