Intended Use - Ahlborn ZED7 00-ABx Instruction Manual

Measuring module for voltage, current and power
Table of Contents

Advertisement

3 Safety Notes
→ The measuring modul internal current measurement path is not intended for
continuous operation for currents greater than 10A. In that case, the
measurement can only be run for a maximum of 10 minutes. After this
period of time, the device must be cooled down to ambient temperature.
→ Only use the socket adapter for the duration of the actual measurement.

3.1 Intended Use

The fast ALMEMO
for measuring current and voltage signals. These include voltage DC, current DC,
power DC, voltage AC, current AC, and frequency. The active power including the
power factor will be calculated on the basis of these signals. The measuring
module is suitable for voltage measurements up to 400V (DC) and 400V
well as for current measurements up to 20A (DC) respectively 20A
The socket adapter is used as an aid for measuring voltage, current and power on
circuits that are electrically connected directly to the low-voltage mains (CAT II).
The socket adapter is to be used exclusively for operation with the measuring
module. Other modes of operation are not permitted.
- 6 –
®
measuring module for voltage, current, and power is intended
(AC) as
eff
(AC).
eff

Advertisement

Table of Contents
loading

Table of Contents