5 How to run a Measurement with the Measuring Module
To avoid electric shocks, do not touch any exposed parts or
connections during operation.
Use the supplied touch proof connector cables.
With no consumer present do not connect the current measurement
module directly to the voltage source. Possible resulting risks are
overheating or destruction of the module.
Voltage and current are measured synchronously and the power will
be calculated on the basis of those two.
1. Connect a consumer to the current source.
2. Switch off the consumer to make sure that no current flows and check that
no voltages are present.
3. Connect the ALMEMO
socket of the ALMEMO
4. Switch on the ALMEMO
5. Integrate the measuring module in the measuring path of the consumer via
the blue connecting cable (into socket A).
6. Connect the black connecting cable to the black socket (COM) of the
measuring module and directly to the voltage source. It should be noted that
the middle socket (COM) represents the ground potential of the measuring
module.
7. Using the red connecting cable, connect the red socket (V) directly to the
voltage source.
- 18 –
®
plug of the measuring module to an ALMEMO
®
measuring device.
®
measuring device.
®
input