How To Measure Power Dc With The Module Zed7 07-Abxy - Ahlborn ZED7 00-ABx Instruction Manual

Measuring module for voltage, current and power
Table of Contents

Advertisement

2. Switch off the consumer to make sure that no current flows and check that
no voltages are present.
3. Connect the ALMEMO
socket of the ALMEMO
4. Switch on the ALMEMO
5. Integrate the measuring module in the current path of the consumer
whereby the blue connecting cable must be connected to the positive pole
and the blue socket (A). The black connecting cable must be connected to
the ground terminal (COM).
6. Switch on the consumer.
➢ The measured values will be displayed on the ALMEMO
5.3 How to measure Power DC with the
Module ZED7 07-ABxy
The device internal current measurement path is not intended for
continuous operation for currents greater than 10A. In that case, the
measurement can only be run for a maximum of 10 minutes. After this
period of time, the device must be cooled down to ambient
temperature.
Cable the system (especially for measuring voltages greater than 50V)
in power-off state. Only switch on the voltage after the cabling has
been set up completely.
5 How to run a Measurement with the Measuring Module
plug of the measuring module to an ALMEMO
®
®
measuring device.
®
measuring device.
input
®
®
measuring device.
- 17 –

Advertisement

Table of Contents
loading

Table of Contents