➢ The measured values will be displayed on the ALMEMO
5.2 How to measure current DC with the module ZAD
901-ABx
For currents higher that 10A, the device internal current measurement
path is not intended for continuous operation. A measurement shall
last for no longer than 10 minutes at most. After 10 minutes, the
device must cool down to ambient temperature.
Wire in a zero-potential condition (especially for measurements above
50V). Switch on the voltage afterwards.
During operation never touch any exposed connections and
components to avoid electric shocks.
Use the supplied touch protected connection cables.
Never connect the current measuring module without a load directly to
the voltage source because by doing so the module could be
destroyed and there is danger of overheating.
1. Connect a load to the current source.
2. Switch off the load to make sure no current flows and check that no voltage
is applied to the module.
3. Insert the ALMEMO
socket of the ALMEMO
4. Switch on the ALMEMO
5 How to measure with the measuring module
®
plug of the measuring module into an ALMEMO
®
measuring device.
®
measuring device.
measuring device.
®
®
input