Typical Specifications - Viavi E4123MM User Manual

Optical time domain reflectometer for t-berd®/mts-2000, -4000 v2, -5800, smartotdr, celladvisor 5g and oneadvisor-800 platforms
Table of Contents

Advertisement

Typical specifications

Typical values, measured at 25°C unless specified.
Central
1
Wavelength
Cut Wavelength
range on live
port (filtered)
Typical RMS
2
Dynamic Range
Distance Range
Pulse width
Event Dead
3
Zone
Attenuation
4
Dead Zone
Splitter
Attenuation
Dead Zone
1. Laser, at 25° C and measured at 10 µs
2. Typical value corresponding to the one-way difference (in dB) between the extrapolated backscattering
level at the beginning of the fiber and the RMS (SNR = 1) noise level, after 3 minutes averaging and
high dynamic resolution, using the largest pulse width.
3. EDZ measured at ± 1.5 dB below the peak of a unsaturated reflective event using the shortest pulse
width.
4. ADZ measured at ± 0.5 dB from the linear regression, using a FC/UPC- type reflectance, at shortest
pulse width.
5. At 1650 nm
6. At 1550 nm
100A Series
1310 nm ± 20nm
1550 nm ± 20nm
filtered 1650 nm ± 20 nm
1290 - 1580 nm
Isolation > 45 dB
37 / 35 / 32 dB
5 ns to 20µs
40 m after 12dB splitter
5
loss
User Manual
OTDR Technical specifications for SmartOTDR
100AS Series
1310 nm ± 20nm
1550 nm ± 20nm
30 / 30 dB
Up to 260 km
1.35 m
4 m
NA
700MAN201 Rev. 004
Chapter 5 Technical specifications
100B Series
1310 ± 20nm
1550 ± 20nm
filtered 1625 nm ± 20nm
filtered 1650 nm ± 20nm
E136FB Module
1290 - 1580 nm
Isolation > 45 dB
E138FB65 Module
1260 - 1620 nm
Isolation > 45 dB
40 / 40 / 41 dB
3 ns to 20µs
0.90 m
45 m after 15dB splitter
2.5 m
6