Grouped Events - Viavi E4123MM User Manual

Optical time domain reflectometer for t-berd®/mts-2000, -4000 v2, -5800, smartotdr, celladvisor 5g and oneadvisor-800 platforms
Table of Contents

Advertisement

Chapter 2 SMART TEST function
Results display
The horizontal shift is performed maintaining the point of intersection between the trace
and the selected cursor at the same level, scrolling the trace horizontally while following
it vertically, so that it never goes off the screen.
To use this function:
1
Click on the trace.and displace trace manually on touchscreen toward left/right or
upward/backward.

Grouped events

If two events are very close to each other, they are grouped (see
20
to get a description of the SmartLink view with grouped events).
In Trace view:
1
In the table, the loss of the group will be displayed at 0.0dB.
2
The reflectance of the group will display the worst reflectance of the events in the
group.
3
No loss is displayed on the event level in the group.
4
Estimated event reflectance will be displayed for every event in the group.
Figure 18
Trace results - Grouped events
Table View
From the results screen, click on Table to display exclusively the measurement results
in a table.
26
2
1
3
User Manual
4
700MAN201 Rev. 004
"Grouped events" page

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents