Chapter 1 Starting up
Connecting fiber optic cable
Figure 2
Modules with APC and PC connector
Caution
Never connect a PC connector into an APC test port or vice versa. This will result in
damaging the connector end faces.
Figure 3
PC/APC bad connection
WARNING
VIAVI declines responsibilities of connector damages if a poor quality con-
nector is used or APC to PC connections made. Test port connector repair
will be charged
Connecting Fiber optic cable to test port
After ensuring proper cleaning of both end connectors, follow the below steps in order
to correctly and safely connect the optical fiber into the test port:
4
User Manual
700MAN201 Rev. 004