REPLACING THE DUST FILTER
Notice: if you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page  61.
Proceed as follows:
Open the chamber door.
1
Pull out the dust filter handle from underneath the sterilizer.
2
Lift the used filter from the handle and remove it.
3
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl
Maintenance
69