Diagnostics
Message/Alert
Door Gasket replacement is due.
Have you replaced the door
gasket?
Bact. Filter replacement is due..
Have you replaced the
bacteriological filter?
Dust Filter replacement is due.
Have you replaced the dust
filter?
Device has completed 4000
cycles.
Please call your technical
support to organize service.
Possible leak detected.
Please run Vacuum Test.
Please check:
- Not to overload the sterilizer
- Door gasket
- Dust filter
If the problem persists contact
the service.
Have you filled the clean water
tank?
Please ensure the chamber is
clean and diffuser is present
and well positioned.
86
Description
These messages advise that one consumable must
be replaced.
This message advises you that the target of 4000
cycles has been achieved and the relevant
maintenance step is to be performed.
Air was detected in the chamber: a vacuum leak is
suspected. The cycle was completed but a vacuum
test is required.
This message advises you that the pressure inside
the chamber did not drop as expected in the first 30
seconds of drying phase.
There is not enough water in the tank.
This message advises you that the steam diffuser
plate is missing or isn't correctly positioned.
Replace the consumable and tap
page  60).
If you don't replace the consumable, press
In this case, you may still use the sterilizer but the message will appear again after
some cycles.
CAUTION! Operating the sterilizer with expired consumables could be
dangerous and could damage the sterilizer.
Call technical service for the 4000 cycle maintenance.
Run a vacuum test.
Call for service if an anomaly is detected.
Check the door gasket and the dust filter correct positioning and make sure don't
overload the sterilizer chamber.
If the problem persists, contact your Technical Service.
Fill the water tank as requested.
Check the chamber cleaning and ensure that the steam diffuser is correctly placed and
engaged.
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl