W&H Lyla Instructions For Use Manual page 88

Table of Contents

Advertisement

Diagnostics
RESET THE SAFETY THERMOSTAT
The sterilizer is fitted with a safety thermostat to prevent it from overheating. If the safety thermostat operates because of too high temperatures,
the error 240 or a timeout error is displayed. The thermostat must be reset manually. Proceed as follows:
Switch the sterilizer OFF and remove the mains cable.
1
Wait for the sterilizer to cool down.
2
Open the chamber door.
3
Remove the dust filter and move the sterilizer closer to the shelf edge.
4
Push the reset button of the thermostat switch: a click sound indicates that the
5
thermostat switch has been reset.
88
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl