Troubleshooting

MANAGING ERRORS
If during a sterilization cycle an error occurs do the following:
Wait until the end of the reset phase.
1
CAUTION! Do not switch off the sterilizer during the reset
phase: it takes some minutes to reset the system and reach
safe conditions in the sterilizer chamber.
When the OPEN button appears, tap it to unlock the door.
2
Confirm the opening of the door.
3
Notice: water could be present in the chamber when opening the
door: prevent spilling (e.g., place a towel below the chamber door).
VIEW AND SAVE THE ERROR LOG
On the homepage, tap
1
errors appears.
Tap
to save the list in the USB pen drive.
2
ERROR PAGE
During the sterilization cycle, the sterilizer is continuously monitored
by a control system. If an anomaly is detected, the cycle is aborted
automatically, and the sterilizer starts a reset phase.
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl
>
>
>
: a list of the last
The following page appears:
WARNING MESSAGES
Message
Description
Load not
The load is not sterile.
sterile
WARNING! Do not use items on patients!
Drying
The load might be wet.
interrupted
WARNING! Wet items are for immediate use only!
Diagnostics
Part
Description
A
Current
sterilization cycle
B
Error number, See
"Errors" on
page  82 .
C
Warning
messages.
Open button that
D
appears after the
reset phase had
finished.