400-Cycle Maintenance - W&H Lyla Instructions For Use Manual

Table of Contents

Advertisement

Maintenance

400-cycle maintenance

REPLACING THE BACTERIOLOGICAL FILTER
Notice: if you replace this consumable before the maintenance due date you have to reset the cycle counter. See "Replace the consumable
before the maintenance due date" on page  61.
Proceed as follows:
Open the service door.
1
Unscrew the bacteriological filter by hand (counter-clockwise).
2
Screw on the new bacteriological filter (clockwise) and tighten it snug.
3
68
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl