W&H Lyla Instructions For Use Manual page 98

Table of Contents

Advertisement

Technical data
Minimum duration of the drying phase (set by the user)
Load type
1
: "dental load" according to EN ISO 17665-3.
STERILIZATION CYCLE DATA (B-TYPE CYCLES – AVAILABLE ONLY WITH ACTIVATION CODE)
Sterilization temperature
Sterilization pressure
Duration of the Plateau/Sterilization phase
Minimum duration of the drying phase (set by the user)
ECO mode for 0.5 kg wrapped load (fixed dry)
Load type
98
S Unwrapped 134
S Wrapped 134
7'
Unwrapped items:
Wrapped items:
Solid
n
n
Hollow B (simple
n
n
hollow item)
n
Hinges device
n
n
1
Dental load
n
n
Handpieces
n
n
Sterilization cycles (B-type cycles – available only with Activation Code)
B Universal 134
134 °C (273 °F)
3.03 bar
2.03 bar (g)
4'
25'
Yes (7')
All unwrapped, bagged, single/double wrapped items:
Solid
n
Hollow A (narrow lumen)
n
Hollow B (simple hollow item)
n
Porous
n
Sterilization cycles (S-type cycles)
S Gentle 121
25'
30'
Solid
Hollow B (simple hollow item)
Hinges device
Porous
1
Dental load
Handpieces
B Prion / Extended 134
134 °C (273 °F)
3.03 bar
2.03 bar (g)
18'30"
25'
Yes (7')
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl
S Prion 134
25'
B Universal 121
121 °C (250 °F)
2.04 bar
1.04 bar (g)
20'30"
30'
No

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rin-210-17Rin-210-22

Table of Contents