Sterilization Cycle Phases - W&H Lyla Instructions For Use Manual

Table of Contents

Advertisement

Technical data

Sterilization cycle phases

COMMON LEGEND OF THE STERILIZATION CYCLE PHASES
Following is the description of the sterilization phases.
Code
Description
PHE
Pre-heating of the sterilizer. This phase is not considered a part of the cycle.
HEA
Heating phase (heat the chamber and the load)
PV1 - PV2
Vacuum pulse (removal of air from the sterilizer chamber/load)
PP1 - PP2
Pressure pulse (steam generation)
PPH
Rise to the Plateau/Sterilization phase
PR
Process (Plateau/Sterilization phase)
Vacuum drying
DRY
LEV
Leveling. Pressure inside the sterilization chamber is leveled to the atmospheric pressure.
END
End of the cycle
102
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl